Gasboard-2513.png

Cubic TDLAS O₂ & NDIR CO₂ Dual-Gas Sensor
Gasboard-2513Gasboard-2513

Gasboard-2513 is an O2/CO2 dual gas sensor developed for medical respiratory applications, utilizing TDLAS technology for oxygen measurement and NDIR technology for carbon dioxide measurement within a single sensor module. It delivers ultra-high selectivity, strong resistance to cross-gas and humidity interference, and a fast response time of <200 ms, supporting real-time monitoring under rapidly changing respiratory conditions.
Gasboard-2513 features calibration-free and maintenance-free operation, superior vibration resistance, and long-term measurement stability, ensuring reliable performance across different clinical and operating environments.
With a compact modular design, ultra-low power consumption, and standard digital interfaces, Gasboard-2513 can be easily integrated into ventilators, anesthesia machines, metabolic analyzers, and CPET systems, enabling simultaneous and stable O2 and CO2 monitoring for both steady-state and dynamic respiratory measurements.
Show More

Features

  • Combined NDIR (Non-Dispersive Infrared) and TDLAS (Tunable laser absorption spectroscopy) technologies
  • Accurate O2 and CO2 measurement with ultra-high selectivity
  • Stable long-term performance with extended service life
  • High resistance to vibration and electromagnetic interference
  • Maintenance-free and calibration-free operation
  • Modular design for easy integration into medical equipment

Specifications

CO2 Sensor CM1106-C
Detection Principle O₂: TDLAS (Tunable Diode Laser Absorption Spectroscopy) technology
CO₂: NDIR (Non-Dispersive Infrared) technology
Detection Range O₂: 0–100 %VOL
CO₂: 0–10 %VOL
Accuracy① O₂: ±1 %FS + 2 %reading
CO₂: ±0.1 %Vol @ 0–5 %Vol ±3 %reading @ 5–10 %Vol
Resolution O₂: 0.01 %Vol
CO₂: 0.0001
Warm-up Time O₂: ≤5 min @ 25 °C
CO₂: ≤2 min @ 25 °C
Working Temperature O₂: 0–50 °C
CO₂: 0–50 °C
Working Humidity O₂: 0–98 %RH (non-condensing)
CO₂: 0–98 %RH (non-condensing)
Response Time O₂: 200 ms @ 250 mL/min (T10–T90)
CO₂: 300 ms @ 300 mL/min (T10–T90)
Power Supply O₂: 5 V ±5 %
CO₂: 5 V ±5 %
Operating Current O₂: ≤150 mA (typical 100 mA)
CO₂: Pulse current: <200 mA Average current: <100 mA
Output Interface O₂: TTL (3.3 V)
CO₂: TTL (3.3 V)
Storage Conditions O₂: −40~85 °C; 0~98 %RH (non-condensing)
CO₂: −40~85 °C; 0~98 %RH (non-condensing)
Flow Range O₂: <500 mL/min
CO₂: 100–400 mL/min
Lifetime O₂: ≥10 years
CO₂: ≥10 years
①: Accuracy is guaranteed only after sufficient warm-up.
Show More